Page17FlowDDE  9.17.067
Anycombinationoftheaboveisvalidbyseparatingthepartsbyacomma,e.g.C(X)!P(0)=110,20,GROUP3willread
parameters1to10,20andallparametersingroup3ofchannelX.
5.2.3 Servertopic
InformationabouttheFlowDDEserver,severalsettingsandFLOWBUSconfigurationinformationcanbeobtainedby
clientsthroughtheservertopic.
Theservertopichasseveralitems:
Server!ProductName FlowDDEorFlowDDE2
Server!FlowDDEVersion VersionofFlowDDE
Server!DatabaseVersion VersionofFlowDDEdatabase(FlowDDE.mdb)
Server!FLOWBUSDLLVersionVersionofFlowb32.dll
Server!Status
 StatusofFlowDDE
Server!ComStatus StatusofFlowDDEIdle,Normalpolling,Simulation,ScanningFLOWBUS
Server!OpenCom Communicationstatus/command0=close,1=open
Server!MaxChan MaximumchannelnumberinFLOWBUSconfiguration
Server!MaxNode MaximumnodeaddressinFLOWBUSsystem
Server!MaxPar Maximumparameternumber
inFlowDDEdatabase
Server!SelChan Selectchannelforchangingpollingmode0=allchannels
Server!Polling Pollingstatus/commandforchannelsetbySelChan0=dis,‐1=enabled
Server!PollTime Polltimeintervalforpolledparameters(notrealtime)
Server!PollBatch Numberofchannelstopoll
Server!RTPInterval Polltimeintervalforrealtimechannels
Server!InterfaceChan ChannelnumberofinstrumentwhichcommunicateswithFlowDDE
Server!Interface Communicationinterface,e.g.COM1
Server!Baudrate RS232communicationbaudrate