Risk of overheating animals. 
Laboratory animals are precious creatures. Always handle them with great care. The 
following  factors  might  lead  to  a  temporary  rise  in  body  temperature  during 
measurements due to Specific Absorption Rate (SAR): 
  excessive RF power 
  excessive gradient usage 
or due to too high ambient temperature: 
  too high water temperature of the animal body conditioning unit 
  malfunction of the gradient air cooling 
  Permanently  supervise  the  animal’s  health  status  during  MRI  measurements, 
especially  body  temperature  and  respiration  rate.  Stop  the  measurement 
immediately if physiological parameters leave their normal physiological range. 
  To reduce SAR, increase repetition time (TR), or reduce matrix or resolution of 
the method if applicable. 
  To  reduce  ambient  temperature,  reduce  the  water  temperature  of  the  body 
conditioning  unit  or  use  air  conditioning.  Check  that  the  air  flow  through  the 
gradient fans is not restricted. 
Risk of peripheral nerve stimulation of animals. 
Excessive  gradient  usage  may  lead  to  peripheral  nerve  stimulation  during 
measurements.  
  Permanently  supervise  the  animal’s  health  status  during  MRI  measurements. 
Should the situation arise stop the measurement immediately.