Icon System Manual T02985_02_03 73
MRI System Operation
5.2.5 Positioning
Turn on the animal warming unit (option). Regarding the usage of this option please refer
to the technical documentation.
CAUTION
Risk of personal injury or injury of the animal.
Leaking water may lead to arcing or electric shocks in the RF Coil or animal bed.
Before preparing the animal or sample always make sure that the warm water tubes
are not leaking. In particular, visually inspect if water entered the animal bed or RF
Coil housing. If water was entering the animal bed or RF Coil,
swab the water completely with a dry and soft cloth;
let all parts dry at temperatures ≤ 40°C before starting any measurement.
Turn on the Isofluorane gas supply and make sure, that the scavenging system is on
(option).
WARNING
Risk of personal injury.
When using narcotic gases (medical gases),
pay attention to the local regulations. In particular you must not exceed the
maximum allowed concentration for medical gases in a laboratory environment.
Carry out appropriate compliance with the regulations;
allow for sufficient fresh air and use extraction system for anesthetic gases;
make sure that the scavenging system is turned on (if present).