78 Icon System Manual T02985_02_03
MRI System Operation
5.2.6.1 Workflow
The coil can be tuned and matched inside or outside the magnet bore. If possible,
tune/match the coil while it is in its final position inside the magnet.
1. Connect the RF Coil at the corresponding socket of the RF interface
2. Start the tune/match from ParaVision (see corresponding chapter in ParaVision
software manual).
3. Adjust and optimize the tune and match condition by turning the tune and match
knobs of the RF Coil (labeled T or M).
4. Stop the tune/match on the RF interface display.