Icon System Manual T02985_02_03 93
Maintenance
8 Maintenance
8.1 Regular Checks
Please read the following instructions carefully and carry out the following regular checks.
8.1.1 Checks before Every Measurement
Check before every measurement:
The gradient cooling fans must not be covered
Check for leaks in the anesthesia system
Visually inspect the RF Coil and its input lead for damage.
WARNING
Risk of fire or serious injury.
The gradient coils may overheat and result in fire or personal injury if the cooling air
flow is constrained. Visually inspect both (left and right) gradient cooling fans at the
magnet before starting the measurement.
Make sure that nothing covers the fan openings.