Icon System Manual T02985_02_03 95
Maintenance
8.2 Cleaning
The following table provides cleaning aspects for different components.
Cleaning the monitor
Clean the monitor using a microfiber cloth. In case the monitor
is contaminated, use a window cleaner.
Immediately remove water drops on the monitor
Avoid impacts and scratches to the monitor
Cleaning the system
covers
Read the safety chapter first.
Clean the covers using liquid household agent
Cleaning the animal
beds
Do not use cleaners or disinfectants containing alcohol, ether or
acetone.
Use a dampened cloth for cleaning
Disinfect the animal bed with a commercially available
disinfecting agent and follow the manufacturer’s
instructions
Cleaning the RF
Coils and connectors
Read the safety chapter first.
Do not use organic solvents, such as alcohol or acetone
Do not submerge the coil or plug in cleaning liquid
Dampen a soft cloth with water or a diluted household
cleaner solution
Carefully wipe the plugs and connectors with the cloth.
Do not touch the contacts
Do not use hard or sharp objects (e.g. knives or
tweezers)