Troubleshooting
Z33119_1_003 41
No Problem Status LED Remote
LED
Cause Solution
7 BCU not visible
in TopSpin in
spite the BCU is
connected with
BSVT
irrelevant Green TopSpin failure Plug out and then plug in the communication cable.
Reboot edte or TopSpin.
8 BCU not visible
in TopSpin in
spite the BCU is
connected with
BSVT
irrelevant Off Communication cable
broken
Exchange the communication cable (Z117511).
9 BCU does not
respond to
command of
TopSpin
irrelevant Green Rotary switch in a manual
mode
Turn rotary switch to position remote.
10 BCU does not
start in spite
rotary switch is
on remote and
sample heater is
on (BVT)
Off Red No switch signal (5V) from
the BVT
Plug in the communication cable from the BVT into the D-
Sub 15 socket of the BCU.
Exchange the communication cable (Z118350)
11 BCU shuts off
after some time
of operation
Red irrelevant BCU overheats Check the wall clearance (min 30 cm) to the inlet and / or
outlet grid of the BCU.
Clean the dust filter inside the BCU (see Maintenance)
Check if the fan is working -> Repair kit for the BCU I:
Z129903 (230 V). Z129904 (115 V).
Service case: contact service.
12 Noise and / or
vibrations from
BCU
irrelevant irrelevant Transport lock has not been
removed.
Fan defective (BCU I)
Transport damage
Unscrew and remove transport lock.
Repair kit for BCU I: Z129903 (230 V). Z129904 (115 V).
Service case: contact service.