Preparation
Z33119_2_003 25
5 Preparation
CAUTION
Accident and material damage hazard due to flying objects
When the operator pressurizes the transfer line before the protective cover is taken off, it can
shoot away.
u Remove the white protective cover of the transfer line before pressurizing the BCU with
VT gas (max. 6 bar).
NOTICE
Instability and poor Vibration Damping of the Unit during Operation
Material damage may occur when the unit is not vibration damped.
u Release and remove the transport safety lock on the bottom of the housing.
u Store the packaging and the transport safety lock. Use only the original packaging for
transportation or shipping.
u Do not place any object on top of the unit.
NOTICE
Material damage hazard from unit overheating.
Material damage may occur when the unit overheats as a result of a blocked air inlet and/or
outlet.
u The unit must be mounted on a level and even surface.
u Around the unit there must be at least a 30 cm deep area clear of all obstructions.
NOTICE
Material damage hazard of the unit due to incorrect handling
Material damage may occur when the unit is repositioned be pulling on the transfer line.
u Do not pull on the transfer line in order to position the unit.