39 Operation and Adjustment Manual
Forinformationonthecali
brationofthe190typemi
crometers,pleasereferto
themanualforyourspe
cificproduct.
2.Pointthetelescopeatanarbitrarilyplacedscale,heldatafinite
distance(werecommendabout4feetaway).Placethescaleina
horizontalorientationandchoosesomearbitrarytargetpoint,e.g.,
thethreeinchtarget.Tightenthehorizontalandverticaltangent
locks,andusethetangentscrewstosetpreciselyonthetarget.
3.Withoutdisturbingthemicrometerdrum,gentlyrotatethemi
crometerbodyonthetelescopebarrel180°.Nowthedétenteposi
tiononthebottomofthemicrometershouldbealignedwiththe
pinonthetelescopebarrel,althoughagainnotseated.
4.Nowlookthroughthetelescopetoseeifthereticleisstillcen
teredonthechosenscaletarget.Ifithasmoved,rotatethemi
crometerdrumsothatthetargetcomesbackintoregisterdirectlyin
thecenterofthereticleasitwasbefore.
5.Readthemicrometerdial.Ifithasdeviatedfromzeroineither
directionbymorethan0.001”,themicrometerisoutofadjustment
andwillintroduceerrorintothefieldchecksdescribedinthischap
ter.Pleaserefertothemicrometermanualforinformationregard
inghowtomaketheappropriateadjustments.Ontheotherhand,
ifthemicrometeriswithintheacceptable±0.001”band(indicating
anactualerrorofhalfthereading),thenyoumayproceed.
Tocheckforrangeadjustment:
1.Installandseatthemicrometeronthetelescopebarrel.Setthe
micrometerdrumtozero.
2.Pointthetransitonceagaintoanarbitrarilychosentargetloca
tion(linepair)onahorizontallymountedscale.Besurethatthe
scaleisnominallysquaretothetransit.
3.
Usingthemicrometeronly,shiftthereticleovertothenextadja
centlinepair,representingashiftofeither0.100”(Englishscale)or
2mm(metricscale).Notethemicrometerreadingatthispoint.
4.Againusingthemicrometeronly,shiftthereticleintheopposite
directionbackpastthestartingpoint,tothenextadjacentlinepair.
Notethemicrometerreadingatthispointaswell.
5.Iftheerrorineitherpositionisgreaterthan±0.0005”(English
scale)or0.01mm(metricscale),themicrometershouldbecali
bratedpriortouse.Pleaserefertothemicrometer
manualorreturn
tothefactoryforadjustment.
Therangeadjustmentfield
checkdescribedhereisonly
validformicrometershav
ingarangeof±0.100”or
±2.5mm.
Indescribingtherange
check,weareassuming
thatyouareusingametric
scaleinconjunctionwitha
metricmicrometer;orlike
wise,anEnglishscalein
conjunctionwithanEng
lishmicrometer.