EasyManua.ls Logo

Burkert 8045 - About the Manual; Symbols Used; Definition of the word device

Burkert 8045
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5
8045
ENGLISH
QUICKSTART
To access the CALIBRATION MENU simultaneously press for 5 seconds.
l Ensure that the enter key is unlocked - section 4.1.
l Reference can additionally be made to the menu guide - section 4.2.
2.2 PROGRAMMING
The actions which are highlighted in grey must be fully completed for
accurate measurement.
Choose the
desired language
Select the desired
units
Enter the K Factor
Select current
output range
Select the pulse
output parameters
Configure relays
1 & 2 (if available)
Choose the
desired filter
Reset the totalizer
Selection of the
AC Frequency
Measurement or Simulation (optional)
ENTER
See section 4.4.1
See section 4.4.2
See section 4.4.3
See section 4.4.4
See section 4.4.5
See section 4.4.6
See section 4.4.7
See section 4.4.8
See section 4.4.9
Select the units l/s with 1 decimal
point and select m
3
for the totalizer
display.
K
8045
= K factor of the fitting x
F
s
coeff. on the sticker.
4 mA = 0 l/s and
20 mA = 10 l/s.
1 pulse corresponds to 100 l.
2 types of filter available each with
10 levels of damping.
To avoid resetting of the totalizer,
set the internal switch 2 into the
locked position.
To cancel any noise generated
by the main AC frequency.
See the next page.
2.2

Other manuals for Burkert 8045

Related product manuals