CalibratingthepHsensorwithoutabuffersolution:
TheChlorinegeneratorallowsyoutocarryoutaquickcalibrationwithoutusingabuffersolutiontocorrect
smalldeviations,withouttaking thesensor outof itshousing, bycomparing thevalue returnedby the
sensorwiththatreadbyanothermeasurementdevice(thathasbeenrecentlycalibrated).
Currentmeasurement=valuemeasuredbythesensor
Targetmeasurement=adjustthisvalueusingthePlus/Minuskeysuntilitmatchesthecurrentmeasurement.
OncetheTargetMeasurementhasbeenadjusted,pressOK,thesensorwillaligntothisvalue.
CalibrationoftheRedoxsensorusingbuffersolutions:
Usethe465mVbuffersolutiondeliveredwiththeRedoxkitandthe"neutral"rinsingsolution,andfollow
thestepsshownonscreen.
Oncethetargetvaluehasbeenadjustedto465mV,pressOK.Thesensorwillaligntothisvalue.
CalibrationoftheRedoxsensorwithoutusingbuffersolutions
TheChlorinegeneratorallowsyoutocarryoutaquickcalibrationwithoutusingabuffersolutiontocorrect
smalldeviations,withouttaking thesensor outof itshousing, bycomparing thevalue returnedby the
sensorwiththatreadbyanothermeasurementdevice(thathasbeenrecentlycalibrated).
Currentmeasurement=valuemeasuredbythesensor
Targetmeasurement=adjustthisvalueusingthePlus/Minuskeysuntilitmatchesthecurrentmeasurement.
OncetheTargetMeasurementhasbeenadjusted,pressOK,thesensorwillaligntothisvalue.
Calibrationofthetemperaturesensor
Thetemperaturesensoriscalibratedbyadjustingthevalueindicatedbythedevicetomatchthatofa
thermometerconrmedtobeexactandprecise.
Currentmeasurement=valuemeasuredbythesensor
Targetmeasurement=adjustthisvalueusingthePlus/Minuskeysuntilitmatchesthecurrentmeasurement.
OncetheTargetMeasurementhasbeenadjusted,pressOK,thesensorwillaligntothisvalue.