EasyManua.ls Logo

CAIRE Hi Flow Stroller - Troubleshooting Chart; Troubleshooting Charts

CAIRE Hi Flow Stroller
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Hi Flow Stroller Technical Service Manual • PN 14314465 Rev F
17
Hi Flow Stroller
Troubleshooting Chart
X
Symptom Probable Cause Corrective Action
1) Unable to start fill or
excessivly long fill times
a) QDV not properly engaged on
the reservoir
Make sure the QDV on the portable and reservoir are properly
aligned. Ensure that a downward force is being applied to the
portable for top fill. Ensure the portable is completely engaged
for side fill.
b) Reservoir is empty Swap or re-fill the reservoir
c) Vent valve not open Ensure that the vent valve lever is fully in the open position.
The lever must be open to begin a fill.
d) FCV is open Be sure that the FCV knob is in the off (“0”) position. If the
valve is open, fill times can increase.
e) Reservoir saturation pressure is
too low
Swap reservoirs or allow the reservoir time to stabilize and build
pressure
f) Vent valve is obstructed Inspect the vent tubes for blockages. Clean by blowing out with
compressed gas or replace parts if necessary.
g) Leak in the system Check the portable for leaks (RP18) and repair if needed.
h) QDV damaged or faulty Inspect the QDV and be sure the poppet opens properly and
smoothly. If necessary, replace the QDV (RP15 or RP16)
i) Faulty vent valve Replace the vent valve (RP19)
2) Liquid leaks from the
coupled QDVs during
the fill
a) Worn or damaged lip seal Replace the QDV lip seal if top fill (RP16). See reservoir techni-
cal manual for replacement of side fill lip seal on reservoir.
3) Unable to disconnect
the portable from the
reservoir
a) Pop-off assembly not being uti-
lized (Top fill only)
Ensure that the pop-off assembly on the reservoir is being used
(top fill only). Do not use force to separate the QDVs on top fill
or side fill.
b) QDVs are frozen together Leave the units coupled with the vent valve closed and let them
sit until they warm up enough to disconnect. Always ensure
that male and female QDV’s are cleaned and dried prior to each
fill.
4) Liquid leaks from the
QDV poppet after fill-
ing
a) Ice crystal preveting the QDV
from closing properly.
Engage and disengage the portable onto the reservoir several
times to dislodge the ice crystal. Always be sure that the male
and female QDVs are wiped clean and dry before filling.
b) Dirty or damaged QDV poppet Replace the QDV (RP15 or RP16)
5) Liquid leaks from the
vent valve tube/outlet
a) Vent valve is not fully closed Ensure that the vent valve lever is fully in the closed position.
b) The portable has been trans-
ported or laid in an improper
operating position
Return the portable to an upright or acceptable operating posi-
tion and allow several minutes for stabilization.
c) Vent valve is frozen open Allow the portable to warm until the vent valve can close. After
the warm up, allow up to 60 minutes for the portable to stabilize
and build pressure before operating.
d) Faulty vent valve Replace the vent valve (RP19)

Related product manuals