6. Specifications
Table 6-1: Measurement specifications
Measurement Range Resolution Accuracy or repeatability
Air temperature –50 to 60 °C 0.1 °C ±0.6 °C
Vapor pressure 0 to 47 kPa 0.01 kPa
Varies with temperature and humidity,
±0.2kPa typical below 40 °C.
FIGURE 6-1 (p. 11) shows accuracies for
different temperature and humidity values.
Relative
humidity
0 to 100% 0.1
Varies with temperature and humidity,
±3%RH typical. FIGURE 6-2 (p. 11) shows
accuracies for different temperature
and humidity.
Barometric
pressure
500to1100hPa 0.1 hPa
±1 hPa (–10 to 50 °C);
±5 hPa (–40 to 60 °C)
Wind speed
0 to 30 m s
1
0.01 m s
1
0.3 m s
1
or 3%, whichever is greater
Wind speed
max 10 s gust
0 to 30 m s
1
0.1 m s
1
0.3 m s
1
or 3%, whichever is greater
Wind direction 0 to 359° ±5°
Solar radiation
0to1750Wm
2
1 W m
2
Accuracy: ±5% of measurement typical;
Stability: <2% per year
Precipitation
0 to 400 mm hr
1
0.017 mm
±5% of measurement from
0 to 50 mm hr
1
Tilt ±90 0. ±
Lightning strike
count
0 to 65535 strikes 1 strike
variable with distance,
>25% detection at <10 km, typical
Lightning
average
distance
0 to 40 km 3 km variable