range, drops from the flared hole can miss the gold electrodes entirely. Use a torpedo level,
the bubble level, or the internal tilt measurements to confirm that the ClimaVUE 50 is level.
Symptom: No wind speed (–9,990, –9,999 or –7999)
1. Check anemometer pathway to ensure debris is not blocking the path of the sonic
transducer measurement (between transducers and acoustic mirror on base).
2. Check the sonic transducers for water build-up. Use a dry cloth to remove moisture.
3. Check to see that the sintered glass plate (FIGURE 9-3 (p. 32)) is not dirty.
4. Clean by flushing with water and dry with a dry cloth.
5. Ensure that the ClimaVUE 50 is level.
Symptom: No temperature reading
1. Check the temperature needle to be sure it is not pushed in, which will break the
thermistor wires.
CAUTION:
Always gently handle the temperature sensor needle. It has delicate wires that can be
easily damaged.
ClimaVUE 50 Compact Digital Weather Sensor 36