EasyManua.ls Logo

Canberra 2016 - Performance Adjustments; Pole;Zero Matching Using an Ultra Lege Detector and Fe

Default Icon
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B. Performance Adjustments
At high count rates, the pole/zero (P/Z) matching adjustment (compensation) is ex
-
tremely critical for maintaining good resolution and low peak shift. For a precise and
optimum setting of P/Z matching, a scope vertical sensitivity of 50 mV/div should be
used.
Higher scope sensitivities can also be used, but result in a less precise P/Z matching
adjustment. However, most scopes will overload for a 10 V input signal when the ver-
tical signal returns to the baseline. Thus the P/Z matching will be incorrectly adjusted
resulting in a loss of resolution at high count rates.
When performing the following manual P/Z matching adjustments, set the scope’s ver
-
tical sensitivity to 50 mV/div and use a clamp box, such as the Canberra Model
LB1502 Schottky Clamp Box, to view the unipolar signal. This will eliminate potential
scope overload, and allow precise manual P/Z matching adjustment.
Pole/Zero Matching Using an Ultra LEGe Detector and
55
Fe.
1. Set the gain for a Unipolar Output signal of 8 to 10 volts.
2. Adjust the
55
Fe source count rate to be between 2 kcps and 25 kcps.
3. Using a clamp box, such as the Canberra Model LB1502 Schottky Clamp
Box, observe the AMP OUT signal on the scope and adjust the manual P/Z
control so that the trailing edge of the unipolar pulse returns to the baseline
with no overshoots or undershoots.
37
Pole/Zero Matching Using an Ultra LEGe Detector and
55
Fe.

Table of Contents