CANNON Instrument Company | Unpack & Assemble
17
Warning: Monitor the bath liquid level closely when operating the CT-2000 at
higher temperatures (100 °C to 150 °C). The bath liquid expands as the
temperature increases. The CT-2000 bath jar is not designed to contain liquid
under pressure. If the bath is overfilled the liquid may overflow and damage
equipment.
Drain the Bath
If it becomes necessary to drain the liquid from the bath, follow the steps outlined in this section.
1. Obtain a suitable container to hold all of the liquid drained from the bath (approximately 20 to
22 liters or 4.5 to 5 gallons), ensuring that the bath liquid is within 10 °C of ambient
temperature.
2. Insert a tube into the bath chamber from the top opening and siphon the liquid from the bath
into a container positioned lower than the bath.
Warning: Always use a rubber bulb or similar device to apply suction to a tube
containing bath liquids.