26
CT-2000 Instruction Manual
5. A new calibration constant has been stored for that test temperature.
6. You may now recalibrate for any test temperature at any time.
Calibration Theory
The CT-2000 stores user calibration information in a “bin” corresponding to every temperature
integer. This calibration adjustment value for the bin is effective at all temperature ranges within 0.5
degrees of the temperature integer.
The CT-2000 combines user calibration input with readings obtained by the temperature probe and
data from the general bath offset, a protected menu option, which is preset at the factory by
Cannon Instrument Company (see the MENU Options section for more information).