Command
Type ($$)
Argument
(&..&)
Action Description
displays. If the transfer fails, EE_NO displays. Or ERR.
For example, if this command is sent to a CT-2000 with an address
of 8: /8CEE; the response may be -8CEEYES
Transmit all the whole degree temperature Bins for either °C or °F
or ERR is displayed. This response has the Bin number in the first
five columns followed by two spaces, and then the sign (if any, a
space is none) and the offset. For example, if the following query is
sent to a CT-2000 at an address of 9: /9QTBC this queries the °C
Bin. The following report is received: 10C 0.000; 11C +0.123
(continues in order with….)
149C -0.456
150C -0.789
-9QTBYES
(end of report identifier)
Transmit a report of key bath parameters. This is a single line
report with a network address in the first column followed by the
time elapsed since the last affected temperature: One space and
then a C or F to designate the temperature scale, followed by at
least one space and then one or two digits to indicate hours
followed by a colon, two digits for the minutes, a decimal and two
digits for the seconds. This is followed by at least one space and
then the desired temperature (to the thousandths of a degree).
For example, if the following query is sent to a CT-2000 in °C with a
target temperature of 40 °C and at address 3: /3QTR. This queries
the temperature report. The following report is the response:
3 C 12:34.56 40.000 40.002 -3QTRYES [end of report identifier]
Query the machine type and firmware version. A single line report
is issued indicating the model, number designation and the
firmware revision level for the instrument at this address.
For example, if the following query is sent to a CT-2000 at network
address 7: /7QMT; this queries the machine type. The following
report is received: -7QMTMTCT2000 FV01.01