EasyManua.ls Logo

Canon Omnera 500A - Radiation Protection

Canon Omnera 500A
186 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OMNERA® 1000-095-071-EN Rev. 1.1
Operator´s Manual
Page 24 2020-09-14
2.10.1 Radiation protection
Because of the ionizing nature of x-ray radiation, precautions have to be taken to minimize
the harmful effects to patients and operators/staff during exposures. The aim is to achieve
dose levels “as low as reasonable achievable”. National regulatory dose limitation require-
ments have to be followed.
Following four main factors control the amount (dose) of radiation received from a source:
Patient and operator dose:
Loading factors: Reducing the loading factors reduces the effective dose proportionally.
Lower values will give more noise in the image.
Distance: Increasing the distance reduces dose levels according to the inverse square law.
Beam size: Keep the beam size as small as possible.
Shielding: Whenever possible/necessary protective shielding should be used to limit dose
levels.
2.10.1.1 Protection against primary radiation (patient)
Following measures should to be taken to limit patient dose.
Observe national dose limit regulations.
Exposure parameters (time/mA) should be set as low as possible with an acceptable im-
age noise level.
Set focus to skin distance as large as possible.
Always collimate the exposure field to the area of interest. This will both decrease the dose
level and improve the image quality (less scattered radiation).
If possible/necessary use protective shielding.
2.10.1.2 Protection against secondary radiation
As the patient is the most significant source of scattered radiation during an x-ray exam, the
staff and/or operator will unavoidable be exposed to ionizing radiation when inside the x-ray
room during an exposure. Radiation doses from scattered radiation can be significantly high.
Following safety measures should be taken to minimize scattered radiation to the staff.
Increasing the distance to the central beam reduces dose levels according to the inverse
square law.
Protective clothing, e.g. lead apron, should always be used.
Exposure parameters (time/mA) should be set as low as possible.
Using high kV and low mA produces less scatter.
Always collimate the exposure field to the area of interest.
Added collimator filter reduces the scatter.
Compression of patient.
Safety
Radiation and X-ray tube

Table of Contents