8200-M129-O1 Rev D 205 Installation & Technical 40
SETUP AND CALIBRATION, CONT.
tyPE0= (Continuous 0 Output Format)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER` key again to save it. Otherwise, use the /UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it. Allowable values are:
0 = SMA 1 = SB-400 2 = SB-200
3 = Rice Lake IQ355 4 = AnDFV 5 = WI110
6 = Number Only 7 = Toledo Long/Short 8 = SB500 with Traffic Light
9 = Ranger 5000 Format A
If tYPE0=7 (Toledo Long/Short) is selected, an additional prompt, tShort will
be displayed.
tShort= (Toledo Short)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER` key again to save it. Otherwise, use the /UP ARROW key to
toggle to a new setting and then press the TARE/ENTER key to save it.
tShort= YES tShort= no
Toledo Short Output Format Toledo Long Output Format
SMA – (Type = 0)
If SMA is selected, the data will be transmitted in the following format:
<lf><s><r><n><m><f><xxxxxx.xxx><uuu><cr>
Where:
lf = Line Feed
s = Flags Z= center of Zero, O = Overcap, E = zero Error,
e = weight not currently being displayed
r = Range 1, 2, 3, ...
n = Mode G = Gross, T = Tare, N = Net
m = Motion M = Motion, " "(blank) = no motion
f = Custom Custom flag
xxxxxx.xxx = Weight Six digits with decimal point
uuu = Units ton, lb , l/o, oz , t , kg , g
cr = Carriage Return (hex 0D)
SB-400 – (Type = 1)
If SB-400
or Computer is selected, the data will be transmitted in the following format:
<s><xxxxxx><d><uu><m><cc><cr>
Where:
s = Sign "-" = negative, " " (blank) = positive
xxxxxx.xxx = Weight Six digits
d = Decimal point Added to string if enabled in setup
uu = Units tn, lb, l/o, oz, t, kg, g
m = Mode G = Gross, N = Net
cc = Weight Status OC = overcap
CZ = center of zero
MO = motion
ee = weight not currently being displayed
cr = Carriage Return (hex 0D)
The SB-80, SB-300 (multiple displays not supported) and WinVRS use the SB-400 format.