8555-M512-O1 Rev D 01/15 23
SCALE
With SCALE displayed, press the PRINT/ENTER key. The display will change to CAPACITY.
Proceed to the CAPACITY (Scale Capacity) parameter.
Otherwise, press the ID/HEIGHT key to advance to the ZERO menu.
CAPACITY (Scale Capacity)
The display will show CAPACITY and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys,
enter a new value and then press the PRINT/ENTER key to save it.
Allowable values are 1 through 65,535.
UNITS (Weighing Units)
The display will show UNITS and the current setting. If the value displayed is acceptable,
press the PRINT/ENTER key to save it. Otherwise, using the numeric keys, enter a new
value and then press the PRINT/ENTER key to save it.
Allowable values are 0 through 4.
0 = None 2 = Pounds/Kilograms 4 = Kilograms/Pounds
1 = Pounds Only 3 = Kilograms Only
NOTE: If Weighing Units is set to None or Pounds Only (UNITS = 0 or 1),
the Height Units prompt will not be shown. During operation, the height
units will be determined by the weighing units selection. For example, if the
weight display is in pounds, the height units will be in feet/inches.
INTERVAL (Interval Setting)
The display will show INTERVAL and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys,
enter a new value and then press the PRINT/ENTER key to save it.
Allowable values are 1, 2 or 5.
dECIMAL (Decimal Point Precision)
The display will show dECIMAL and the current setting. If the value displayed is acceptable,
press the PRINT/ENTER key to save it. Otherwise, using the numeric keys, enter a new
value and then press the PRINT/ENTER key to save it.
Allowable values are 0, 1, 2 or 3.
0 = # # # # # # 1 = # # # # #.# 2 = # # # #.# # 3 = # # #.# # #
MOTION RANGE (Motion Range)
The motion range is the number of divisions of change permitted before indicating unstable
(the STABLE annunciator turns off).
The display will show MOTION RANGE and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys,
enter a new value and then press the PRINT/ENTER key to save it.
Allowable values are 0 through 9.