EasyManuals Logo

Cardinal MED VUE Mv1 User Manual

Cardinal MED VUE Mv1
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
8555-M512-O1 Rev D 01/15 48
<LF>P<CR>
Weight analyzer will respond with all information formatted with each item on its own line.
<LF>J<CR>
Weight analyzer will respond with all information formatted with each item separated by
commas (comma delimited) followed by a carriage return.
<LF>Z<CR>
Zero the scale within the constraints of the zero settings.
<LF>T<CR>
Scale attempts to tare itself setting the tare weight to the current gross weight.
<LF>T<xxxxxx.xxx><CR>
Scale attempts to set the weight (represented as<xxxxxx.xxx> in the string) to the
requested value.
<LF>M<CR>
Scale returns the current tare weight.
<LF>U<CR>
Scale will toggle between calibration units if conversion units have been set up in scale
calibration.
<LF>A<CR>
Scale will respond with:
<LF>MedVue<CR>
<LF>XH<CR>
Scale will respond with the current height:
<LF>Height<CR>
<LF>XB<CR>
Scale will respond with the current BMI:
<LF>BMI<CR>
<LF>XI<CR>
Scale will respond with the currently entered ID:
<LF>ID<CR>
<LF>XA<CR>
Scale will respond with current weight, height, BMI, and ID in comma delimited format:
<LF> weight (see <LF>W<CR> above), Height, BMI, ID<CR>
NOTE: If the MedVue weight analyzer has been set to continuous mode of outputting weight,
it will output the weight in the format listed above for <LF>W<CR> every 500 milliseconds.