26
ENG
Gas Leakage Detector +0300047IE rel 1.1 - 26.04.2023
Maintenance
Supply the gas at the known concentration using the
calibration kit, suitably installed as described in the
previous chapter.
Wait about 1 minute until the gas concentration
stabilises.
Click Span Calibration to set the calibration concen-
tration.
Enter the room temperature and relative humidity.
These values will be entered on the calibration certicate to indicate the environmental conditions
during calibration. It is not necessary to use a calibrated instrument to perform this measurement,
an indicative value is sucient.