Gas Leakage Detector +0300047IE rel 1.1 - 26.04.2023
Maintenance
4.3 Calibration via app
Before connecting to the device via the RILEVA TE app, rst make sure that the BLUETOOTH connection and GEOLOCATION are
enabled on the smartphone used.
Make sure that the Bluetooth mode on the GLD small has been activated using the magnetic latch as described in the previous
chapters.
Refer to the Functions chapter of the RILEVA TE app manual for details of all the apps features.
Start calibration on the navigation bar.
Before connecting the calibration kit make sure the sensor is ready and free of gas or other sources
of pollution
Click Fresh air calibration at the bottom right, then select Next
To perform the calibration, the specic gas indicated as the calibration GAS” needs to be used
Enter the gas cylinder reference (serial number of the reference gas or other information to be
shown on the certicate).
Enter the concentration of the gas used for calibration
Click send to device to set the gas concentration used for calibration.