PadScan HD 2
Caresono Technology Co., Ltd 24
Chapter Seven Cleaning and Maintenance
To ensure that the device functions normally, please perform regular cleaning and maintenance of
parts, accessories and probes with mild detergent.
7.1 System cleaning and maintenance
7.1.1 System cleaning
(1)Turn off the device.
(2)Remove the battery from the device.
(3)Use a soft, clean cloth dampened with isopropyl alcohol (or an appropriate hospital cleaning
agent), to clean the device’s surface.
(4)Control the time of sponging according to the instructions of the detergents, and the interval
time should also meet the clinical requirements.
(5)Dry the system’s surface naturally or dry it with clean cloth according to the instructions of
the detergent label.
(6)Use soft and wet cloths with neutral detergents to clean the fingerprint or other filth on the
display screen.
7.1.2 System maintenance
(1)Operate the system in the environment as specified in “1.5”.
(2)After shutting down the device, wait five minutes before restarting the system.
(3)When the device is not used for a long time, pack the device and store in the environment as
specified in “8.2”.
7.2 Probe cleaning and maintenance
Please keep the probe clean to ensure the probe work normally and prolong its service life.