Appendix
A–32 AJ4309 | 2019-08-13
6. When all five images are created, select Generate to generate the gain template.
Defect Correction Template Generation
Before doing this procedure, make sure SID1.2m, no copper is required.
Actual screens may be different from those shown in this procedure.
Note:
Use software post offset mode.
1. On the Defect Calibration page, start the exposure. Eight images need to be
captured.