- 23 -
ERROR MESSAGE & TROUBLE SHOOTING
Errors in weighing mode
Err 01
▶ Cause
The indicator can’t be initialized due to platform shaking.
▶ Management
Put the platform on a flat place without shaking and power on.
Err 02
▶ Cause
The loadcell is connected wrongly or A/D conversion part has a fault.
▶ Management
Check that the loadcell is well connected with main unit.
Err 03
▶ Cause
The calibration isn’t executed or the data of internal memory is lost
due to a electrical shock.
▶ Management
Perform the required specification again in setting mode.
Err 09
▶ Cause
The current weight exceeds zero range.
Err 10
▶ Cause
The specified TARE weight exceeds the maximum capacity.
▶ Management
Set the TARE weight lower than max. capacity.
Or set the maximum capacity higher than the desired TARE weight
in calibration menu. and perform the calibration again.
Err 13
▶ Cause
The initial zero range exceeds above 10% of maximum capacity.
Management
Check the loadcell state.
Over
▶ Cause
Too heavy object is put on platform so that it may exceed the
maximum capacity.
▶ Management
Never put the weight exceeding the maximum capacity
If the loadcell is damaged, replace the loadcell.