- 23 -
ERROR MESSAGE & TROUBLE SHOOTING
Errors in weighing mode
Err 01
Cause
The indicator can’t be initialized due to platform shaking.
Management
Put the platform on a flat place without shaking and power on.
Err 02
Cause
The loadcell is connected wrongly or A/D conversion part has a fault.
Management
Check that the loadcell is well connected with main unit.
Err 03
Cause
The calibration isn’t executed or the data of internal memory is lost
due to a electrical shock.
Management
Perform the required specification again in setting mode.
Err 09
Cause
The current weight exceeds zero range.
Err 10
Cause
The specified TARE weight exceeds the maximum capacity.
Management
Set the TARE weight lower than max. capacity.
Or set the maximum capacity higher than the desired TARE weight
in calibration menu. and perform the calibration again.
Err 13
Cause
The initial zero range exceeds above 10% of maximum capacity.
Management
Check the loadcell state.
Over
Cause
Too heavy object is put on platform so that it may exceed the
maximum capacity.
Management
Never put the weight exceeding the maximum capacity
If the loadcell is damaged, replace the loadcell.