CALIBRATION MODE
Moving method
Open the top cover of indicator and power on with pressing CAL switch in the left
top. At that time, LCD screen shows “ZEro=tESt tArE=CAL” message and pressing
‘TARE’ key moves to CAL 1.
Calibration menu (CAL 1 ~ CAL 5)
CAL 1 : Maximum Capacity Set
CAL 2 : Minimum Division Set
CAL 3 : Setting Weight in Span Calibration
CAL 4 : Zero Calibration
CAL 5 : Span Calibration
Digit input method
NET/GROSS Key : Increase the first place value by 1.
TARE Key : Move to the left by 1 place.
ZERO/MODE Key : Set the inputted value to “0”.
ENTER Key : Move to the next menu.
CAL 1
FUNCTION : Maximum Capacity Set (Range : 1 ~ 99,999)
KEY LC DISPLAY
DESCRIPTION
Input by digit input method.
ENTER : save and go to next menu
CAL 1 condition
200 Kg
CAL 2
FUNCTION : Minimum Division Set (Range : 0.001 ~ 50)
KEY LC DISPLAY
DESCRIPTION
NET/GROSS key:
Input the next division
TARE key:
Input the previous division
ENTER key: Move to the next menu
CAL 2 condition
0.5 Kg
Note 1. The minimum division means the value of one division.
Note 2. External resolution is obtained by dividing the min. division by the
maximum capacity. Set the resolution to be within 1/10,000.
Note 3. The maximum capacity/minimum division is over 10,000, error message
“CH 11” will occur.