Explains the meaning of the 'Warning' symbol indicating severe potential consequences.
Explains the meaning of the 'Attention' symbol indicating potential injury or material damage.
Details connecting load cells and system design considerations for load cell input.
Lists calibration steps including capacity, division, zero, span, and gravity.
Explains how to set the maximum weighing capacity for the scale.
Details setting the minimum division and decimal point for accurate readings.
Configures the number of steps for multi-point calibration.
Explains how to perform zero calibration by emptying the load tray.
Details how to input the known weight for calibration.
Explains how to perform span calibration using a known weight.
Performs zero adjustment when zeroing errors occur.
Details steps for sealing the indicator after calibration using CAL switch and wire.
Describes how to enter the Set Mode using specific key presses.
Explains how to change setup values and navigate menus using keys.
Covers date, time, power off, filters, compensation, and key settings.
Configures device ID, parity, baud rate, usage, and output format for serial communication.
Manages printer selection, print formats, automatic printing, line feed, and item number.
Configures device ID for communication and parity bit settings.
Sets baud rate, usage, output format, and output mode for COM1.
Describes the key sequence to access the test modes.
Lists various tests including key, display, load cell, and communication tests.
Tests the functionality of each key on the indicator.
Checks the LCD and LED displays for proper operation.
Tests load cell connectivity and A/D conversion accuracy.
Verifies serial communication functionality between the indicator and a computer.
Guides on inputting samples for Piece Counting System (PCS) mode.
Guides on directly inputting weight values for PCS mode.
Guides on inputting samples for percent weighing.
Guides on directly inputting weight values for percent weighing.
Lists error codes, causes, and solutions for the weight setup mode.
Lists error codes, causes, and solutions for the weighing mode.