[POWER] key [CAL ] key [O/T] key [UNIT] key [PRINT] key
[Explanatory Operation Sheet]
Key Switch Basic Function
[POWER] Switches between operation and stand-by (On/Off).
[CAL] Enters span calibration or menu item selection.
[O/T] Tares the balance (Displays zero with a container on the pan)
[UNIT] Switches weighing units.
Output the displayed data to an electronic printer or a computer.
[PRINT]
To output date and time → Press and hold for about 3 seconds
Stabilizing Display / Faster Response for Filling
x If it is unavoidable to install the balance at a location subject to air flow or vibration, try
changing the environmental settings to stabilize the reading.
(From mass display) Press [CAL] repeatedly till “E” is blinking in “SEL:EAUS” display → Press
[O/T] (“→E- AbtP8” is displayed with “→” blinking) → Press [CAL] once (“A” blinks) → Press
[CAL] repeatedly until “EA-vibr” is displayed → Press [O/T] (Stability mark is added to “EAvibr”)
→ Press [POWER] repeatedly to return to mass display.
This setting will bring more stability, however response will be slower. If further stability is
desired, select “EA-wind” instead of “EA-vibr” in the above procedure.
x If faster response is required for filling or dosing purpose, select “PoUr” instead of “EA-vibr”
in the above procedure. Pouring mode is set and the response will be very fast, but display
unstable. Refer to 11.2 for details of Pouring mode.
Hints on Fully-automatic Span Calibration(UW series only)
Fully-automatic span calibration PSC
Default setting of PSC is ON. Upon detection of temperature change, span calibration is
automatically performed to maintain accuracy. The Weight symbol blinks prior to
fully-automatic calibration.
How to skip a PSC calibration
Press [POWER] key while Weight symbol is blinking.