12
BAL
Function : Setting the Weight in Span Calibration
KEY LCD SCREEN DESCRIPTION
bAL
ZERO :
Increase the value(1~9)
TARE :
Move the place(value*10)
ON/OFF :
Go to ‘MIN’ menu
HOLD :
Go to below step
500 kg
500
500 kg
HOLD:
Start the Span Calibration
ANY KEY:
Not operate
LoAd
Load the weight which was set in ‘BAL’,
and the press HOLD key.
Note. The weight for span calibration should be within 10~100% of the
maximum capacity in ‘BAL’.
If you operate the Span calibration before checking the value of ‘MIN’/
‘MAX’/‘BAL’, the weight will be not exact in normal mode.
(Do check the vale of ‘MIN’/’MAX’/’BAL with *(STAR) key)