10. ERROR MESSAGE
Err 01
Cause
The Indicator can’t be initialized due to platform shaking
Management
Check the A/D value (at ‘TEST 3’) .I
If the load cell is damaged, replace the load cell
Put the platform on a flat place without shaking and power on.
Err 02
Cause
Initial A/D value of zero is too high or low, when the power is ON
Or in the case of NC-I is lying on the flat
Management
Check the A/D value (at ‘TEST 3’).
If the load cell is damaged, replace the load cell
Check that the load cell is well connected with main unit
Or calibration is done wrongly
Err 09 (Zero Key)
Cause
The current weight exceeds zero range (2 % of Maximum Capacity)
Management
Unload the scale
Err 10 (at Zero calibration)
Cause
Initial A/D value of zero is too high or low.
Management
Check the A/D value (at ‘TEST 3’).
If the load cell is damaged, replace the load cell
Check that the load cell is well connected with main unit
Replace the loadcell.