51
Err 11 (at Span calibration)
Cause
The weight for span calibration is greater than 100% of the maximum capacity.
Management
The weight for span calibration should be within 10~100% of the maximum
capacity in ‘BAL’.
Err 12 (at Span calibration)
Cause
The weight for span calibration is lower than 10% of the maximum capacity.
Management
The weight for span calibration should be within 10~100% of the maximum
capacity in ‘BAL’.
Err 13 (Zero Range when the Power is ON)
Cause
The initial zero range exceeds above 10% of maximum capacity
Management
Unload the scale
Check the A/D value (at ‘TEST 3’).
If the load cell is damaged, replace the load cell
Err 14 (at Span calibration)
Cause
The weight for span calibration is 0kg.
Management
The weight for span calibration should be within 10~100% of the maximum
capacity in ‘BAL’.
BAT or LobAt