12
MANUFACTURER’S DECLARATION
ELECTROMAGNETIC EMISSIONS
The THD2FE is intended for use in the electromagnetic environment (for
home healthcare) specied below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Emissions Test Compliance
Electromagnetic Environment – Guidance
(for home healthcare environment)
RF emissions
CISPR 11
Group 1
The THD2FE uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The THD2FE is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
MANUFACTURER’S DECLARATION
ELECTROMAGNETIC IMMUNITY
The THD2FE is intended for use in the electromagnetic environment (for
home healthcare) specied below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic
Environment –
Guidance
(for home healthcare
environment)
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood,
concrete or ceramic
tile. If oors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Power frequency
(50, 60 Hz)
magnetic eld
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
The THD2FE power
frequency magnetic
elds should be at
levels characteristic
of a typical
location in a typical
home healthcare
environment.