[Note 7] If blood flow rate is slow, lower than 30 ml per minute, and remains
low for 1 minute, audible beeps will sound and the [ALARM] indicator
will be activated. Check donor phlebotomy site to ensure blood flow is
adequate. The beep will repeat once every minute until the flow is
higher than 30 ml per minute.
When the collected volume reaches the preset target volume, the collection
ends.
* The tube clamp closes.
* The tray resumes rocking.
* LED lamp starts flashing.
* The [END] indicator appears to show that the collection is finished.
* The [ML] indicator remains on to show the main display indicates
the net volume of blood in milliliters.
* Beeps sound. The beeps will repeat once every 30 seconds until
[END] button is pressed.
* The [CLAMP] button will not work until [END] button is pressed.
(7) Press [END] button.
* The tray stops rocking.
* The LED lamp stops flashing and remains lit.
* The [GRAM] indicator appears again to indicate the number displayed
is total weight loaded on the tray, represented in grams.
[Note 8] Depressing [TIME] button displays donation draw time on main display.
(8) Crimp blood bag tubing and remove phlebotomy needle from the donor.
(9) Press [CLAMP] button to open clamp and remove blood bags.