19
CM73522UMC.0112
5-3. Audible Alarm Signals
Different audible alarms are generated by a buzzer to indicate that the attention of the
phlebotomist is required.
A. End of normal collection:
----- activated when collection ends normally.
Follow Section 5-2-(7).
----- Three long beeps (1 second long each) each 30 second interval
* Press [
END
] button. Beeps stop and LED lamp stops flashing.
* Crimp blood bag tubing and remove phlebotomy needle.
* Press [CLAMP] button to open clamp and remove blood bags.
B. Slow bleed:
----- activated when blood flow is slower than 30 ml/minute.
Refer to Section 5-2-[Note 7].
----- Five short beeps (once a second, half second long) each 1 minute interval
* Check needling and phlebotomy site to ensure flow is adequate.
* To continue collection, do not press any button. DO NOT press
[END] button as it will clear the volume previously collected.
Remove the reason of slow bleed.
* To end collection, press [END] button.
C. Low battery:
----- activated when battery is exhausted.
Refer to Section 5-4.
----- One very long beep (5 second long) only once
* Recharge the battery, or connect the AC power adaptor.