20
CM73522UMC.0112
D. Tray removal:
----- activated when the tray is removed.
----- Beeps sound twice a second until power is turned off.
* Turn the power switch off.
E. 10 minutes passed:
----- activated when 10 minutes passed after the start of collection.
----- 20 rapid beeps (twice a second) only once for 10 seconds.
* Check needling and phlebotomy site to ensure flow is adequate.
* To continue collection, do not press any button. DO NOT press
[END] button as it will clear the volume previously collected.
* To end collection, press [END] button.
F. 20 minutes passed:
----- activated when 20 minutes passed after the start of collection.
----- Very rapid beeps (four beeps a second) continue.
----- Tube clamp closes.
* Check needling and phlebotomy site to ensure flow is adequate.
* To continue collection, press [CLAMP] button to open the clamp.
* To end collection, press [END] button.
[Remark]
Certain version of the blood collection mixer doesn’t have 10 and 20 minutes alarms
and/or tray removal alarm.