6-3. Leveling the Tray
Tray leveling is ensured by a magnetic position sensor and precision mechanism. After
long use for years, the tray might become unlevel. Leveling of the tray can be simply
adjusted by users.
(1) Prepare a flat (-) tip screwdriver, 3-5 mm (1/8-3/16”) in diameter.
(2) Turn power off and remove tray from tray mount (# 5 in the figure in Section 3-2).
(3) Identify the tray leveling screw (# 8 in the figure in Section 3-1) through the hole.
(4) Turn the screw clockwise, by about a half turn (180 degrees), to make the tray
rotate clockwise. The tray will rotate clockwise by a half degree.
Turn the screw counter-clockwise, by about a half turn (180 degrees), to make the
tray rotate counter-clockwise. The tray will rotate CCW by a half degree.
(5) Replace the tray and turn power on.
(6) Check tray leveling. If tray is not level, repeat from (2) to (5) again.