The circuitry is generally oversized and well ventilated, which avoids
overheating and associated problems.
The recommended operating temperature is between 10 and 30 °C, with
humidity between 30 and 75% and pressure between 700 and 1060 hPa.
When used in a clinic or hospital, in a normal non-sterile environment,
Intelect HPL 7 and HPL15 has no particular protection against accidental
spillage of liquids; occasional splashes or sprays of water or gel do not create
problems due to the distance between circuit boards and ventilation grills.
The recommended environments for use are: hospital, clinic and
private center.
The recommended environments for use have to have the following
environmental and electrical requirements:
-) electrical system that respects the standards of the installation country
-) no reflective surface like mirror
-) laser attention symbol
-) red light outside the laser therapy room
-) labels that underline that it is necessary to use laser protection glasses
during the therapy
-) interlock system connected to the door of the laser therapy room
For the disposal of the machine at the end of its life, follow WEEE
requirements on page 4.