Warning
In the case of power surges caused by direct or indirect lightning strikes, the
machine may lock, preventing laser emission.
In this case contact the Assistance Centre.
Message Probable cause Solution
CHECK MACHINE HOURS
The laser electronics should be
checked
Contact the DJO
Assistance Centre
TEMPERATURE ALARM The working temperature is high
Cool the laser keeping the machine
switched on.
If the problem persists contact the DJO
Assistance Centre
In the case of malfunction the following warnings may be shown on the
Intelect HPL 7 and HPL15 display.
6.3 Alarms