3.12.8 Suppression of automatic stop
Hold the HOLD key pressed and bring the rotary switch
from the OFF position to the position .
The unit emits a double beep then the
P
symbol flashes.
The selected configuration is memorised when the key is
released: the
P
symbol remains lit continuously.
The unit returns to automatic stop mode when the switch
goes to the OFF position.
3.12.9 Activation of the V-Live function
Hold the HOLD key pressed and bring the rotary switch from
the OFF position to the V position.
The unit emits a double beep then the V and symbol
flashes.
The selected configuration is memorised when the key is
released: the V symbol is then ON steady and the symbol
flashes.
Proceed in the same way to suppress the V-Live function: the
symbol then goes OFF when the key is released.
3.12.10 Modification of audio indication threshold in
continuity test
Hold the Yellow key pressed and bring the rotary switch from
the OFF position to the position .
The unit beeps, the
and symbols go ON along with the
threshold value (40.0 by default).
Adjustment is then possible from 1 to 40 by successive
pressings on the yellow key (short pressure: progression of
1 by 1 ; press and hold: progression of 10 by 10 ).
Once the value is chosen, actuate the rotary switch to
memorise.
3.12.11 Default configuration of the unit
Hold the Yellow key pressed and bring the rotary switch from
the OFF position to A position.
The unit emits a double beep then all the segments of the
digital display and the symbol flash.
The default configuration is memorised when the key is
released: the display no longer flashes and the symbol
disappears.
The default configuration is:
- Audio identification threshold: 40
- Auto cut-off: with
- V-Live function: none
3.12.12 Date of last calibration carried out on unit
Hold the MIN/MAX key pressed and bring the rotary switch
from the OFF position to the V position.
The unit emits a beep, then the calibration date of the unit is
displayed in the form “week - year” (SS.AA) as long as the
MIN/MAX key is pressed.