EasyManua.ls Logo

Cheonsei MESTAR+ - Page 12

Default Icon
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
9-5 Unit Setting
unit is Unit setting menu.
Present set unit is displayed, while you enter the Setting
display.
You can change the unit by using the
and
keys.
Set the unit of the main screen and then that of the
secondary screen.
Three types of unit(ppm, mg/L, and %) are provided on
the main screen.
Three types of unit(℃, %, ℉) are supported on the
secondary screen.
When the main screen has the unit shown to be “%”,
the secondary screen can't have the “%” unit (unable to
set the same unit).
When the unit is changed, such settings as dead band,
offset, damping value, and temperature offset will be
initialized.
9-6 Dead band setting
d.bnd is Dead band setting menu.
Present set value is displayed, while you enter the
Setting display.
Dead band value can be changed by
,
Key.
Finish the setting by pressing the ENTER key.
The band can be set up to a maximum of 1.00ppm
(0.30%). The alarm will occur as shown below
example.
Ex) When setting of 0.10ppm
When Setting value of HIGH Alarm is 10.00ppm.
: If measured value is 10.10ppm or over, HIGH
Alarm will occur, and, if measured value become
less than 9.90ppm, HIGH Alarm will be off.
When Setting value of LOW Alarm is 1.00ppm.
: If measured value is less than 0.90ppm, LOW
Alarm will occur, if measured value become
1.10ppm or over, LOW Alarm will be off.
Dead band setting value can not be set within the difference range of HIGH Alarm and LOW Alarm. (Refer
to " 9-4 Alarm Setting" )
Ex) When 1.50ppm HIGH Alarm and 1.00ppm of LOW Alarm are set, 0.50ppm or higher can not be set as
Dead band.