Programmable AC Source 61601/61602/61603/61604 User’s Manual
[SOURce :] CURRent : INRUsh : INTerval
Description : This command sets the measuring interval of the inrush current
measurement.
Query Syntax : [SOURce :] CURRent : INRUsh : INTerval?
Parameters : <NR2>, valid range: 0.0 ~ 999.9 (unit: ms)
Return Parameters : <NR2>
[SOURce :] FREQuency [: {CW | IMMediate}]
Description : The command sets the frequency of the output waveform of the
AC source in Hz.
Query Syntax : [SOURce :] FREQuency [: {CW | IMMediate}]?
Parameters : <NR2>, valid range: 15.00 ~ 1000.00 (unit: Hz)
Return Parameters : <NR2>
[SOURce :] VOLTage [: LEVel][: IMMediate][: AMPLitude] : AC
Description : This command sets the AC composition of output voltage in
Volts.
Query Syntax : [SOURce :] VOLTage [: LEVel][: IMMediate][: AMPLitude] : AC?
Parameters : <NR2>, valid range: 0.0 ~ 150.0 (in low range), 0.0 ~ 300.0 (in
high range)
Return Parameters : <NR2>
[SOURce :] VOLTage [: LEVel][: IMMediate][: AMPLitude] : DC
Description : This command sets the DC composition of output voltage in
Volts.
Query Syntax : [SOURce :] VOLTage [: LEVel][: IMMediate][: AMPLitude] : DC?
Parameters : <NR2>, valid range: -212.1 ~ 212.1 (in low range), -424.2 ~
424.2 (in high range)
Return Parameters : <NR2>
[SOURce :] VOLTage : LIMit : AC
Description : This command sets the setting of Vac LIMIT which will restrict
the value of Vac.
Query Syntax : [SOURce :] VOLTage : LIMit : AC?
Parameters : <NR2>, valid range: 0.0 ~ 300.0 (unit: V)
Return Parameters : <NR2>
[SOURce :] VOLTage : LIMit : DC : PLUS
Description : This command sets the setting of Vdc LIMIT(+) which will
restrict the value of Vdc.
Query Syntax : [SOURce :] VOLTage : LIMit : DC : PLUS?
Parameters : <NR2>, valid range: 0.0 ~ 414.2 (unit: V)
Return Parameters : <NR2>
[SOURce :] VOLTage : LIMit : DC : MINus
Description : This command sets the setting of Vdc LIMIT(-) which will restrict
the value of Vdc.
Query Syntax : [SOURce :] VOLTage : LIMit : DC : MINus?
Parameters : <NR2>, valid range: 0.0 ~ -414.2 (unit: V)
Return Parameters : <NR2>
[SOURce :] VOLTage : RANGe
Description : This command sets output voltage range with three options of
LOW(150 V), HIGH(300 V), or AUTO.