Loading...
- 21 -
Draw. 9. Password-activation the function
Draw. 9-1. Menu password protection
- Administrator
line to write administrator who has access to all set up
- User
line to write user password. User who has access to setting with NO attribute (are not
protected by password)
- Start up
If it is settled on YES during start the balance up user must write access password
(administrator or user)
- Functions
If it is settled on NO (not protected by password) user can use implemented functions
in the balance.
- Set up
If it is set up on NO (not protected by password) user can change setting in the
balance
- Only Kal+GLP
If it is set up on YES user can perform the calibration and calibration report
- Stars
If it is settled on YES during start the balance up password is hidden under starts
Administrator password
Write the password for administrator (max 6 digits) and user. Each admistrator has
access to all functions in the balance. User has access to balance possibilities in
accordance with above description. Please, remember the password. If you set YES
for „Start up” function the password must be writen after switch the balance on.
If the password is not correct using the balance is not possible.
To write the password in use keys described on the drawing 2. or PS/2 keyboard (it
can be connected to the balance port).
Set up the attributes for other options dependly on authorizations for user.