- 23 -
8.1. Automatic balance calibration
It is performed w
hen:
- Period of time passes from last calibration
- temperature changes for settled value by manufacturer
* for balances CX is 0,8° C,
Following information appears on the display:
Draw. 11. Automatic calibration – display view
Time delay in starting the calibration up enables user to take load off the pan untill the
measurements are performed. If the T/O key is pressed the calibration process is stopped.
Set up automatic calibration
Draw. 12. Set up automatic balance calibration
1 – main menu number
2 – function selection factor
3 – function name
4 – name of actual activity
5 – selects factor to autocalibration (time / temp.)
6 – declaring autocalibration time
7 – value of factors for autocalibration
8 – value of time for autocalibration
If the value for autocalibration factor and time changes descriptions for these lines changes
as well (on the drawing field No9. And No 10.)