F none No Function
Unit 1 gm
Unit 2 ct
Unit 3 gm
Stb 2 balance environment set to Normal
CAL Et CAL External
CAL t CAL TEST OFF
Azt ON Auto Zero Tracking set ON
bd9600 Transmission rate
Pr None Parity set to none
Stpbt 1 Stop bit one
Print Req
GLP OFF
A. oFF - no automatic standby
5.2.14 Reset of the balance setting
Reset balance setting and functions to factory setting (rESEt)
ð Select “rESEt” and Press <TOGGLE> key breifly, display show “YES”
ð Press <TARE> key breifly, display show “stored”
The balance is now reset to the factory setting and returns to the weighing mode._
Factory Setting
42
5.2.12 GLP Menu Setting
GLP off : If the user select this option then the balance print format are not
compliance to ISO/GLP/GMP.
GLP on : If the user select this option then the balance print format are compliance to
ISO/GLP/GMP.
Note
If user selected GLP ON do ensure that user print footer for entering into next
transaction and enter into user Menu or Calibration.
5.2.13 A. Off - Setting automatic standby
The automatic standby appreciably extends the operating life of your Battery (If
Install) (Optional)
The balance will enter stand by mode if A-OFF is activated.
The display on the balance remains zero for a specific time as selected in the A.OFF
menu.
A. Off - : no automatic standby (factory setting)
A. Off 1 : automatic standby after 1 minutes
A. Off 5 : automatic standby after 5 minutes
A. Off 10 : automatic standby after 10 minutes
41