09
P-01 pH Standard Solution
After power-up, the unit will enter the measurement mode. Factory default is pH measuring
mode. You can switch between the pH measurement mode and the ORP measurement mode
by pressing the Key. Press to enter set up step. Press to enter electrode
calibration mode.
In P-01, press to enter. Press to select among B1,B2,B3,B4 buffer kinds.
Press to confirm.
Please refer to above description steps to set up P-01.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
B1: 1.68, 4.01, 7.00, 10.01 (US)
B2: 2.00, 4.01, 7.00, 9.21, 11.00 (EU)
B3: 1.68, 4.00, 6.86, 9.18, 12.46 (CN)
B4: 1.68,4.01, 6.86, 9.18 (JP)
5.4 Set UP
5.4.1 pH set up