IMPORTANT: The exhalation valve leaf may stick after cleaning. Before use, check that the exhalation valve leaf operates freely
by gently lifting the valve.
Ensure the valve cover on the Half Mask is in place before use.
WARNING: NEVER DRY THE MASK OR EXHALATION VALVE WITH A CLOTH THAT MAY LEAVE BEHIND LINT.
Lint contamination of the exhalation valve may cause it to leak, resulting in overexposure to contaminants and lead to
sickness or death.
CLEANING THE POWER UNIT
WARNING: The Power Unit contains a battery, sensitive electronics and a motor. DO NOT IMMERSE THE POWER UNIT IN
WATER or use anything wetter than a damp cloth to clean it.
•
After removing the mask, remove the neck support. Leave the filter in place to prevent any dust or liquids from entering
the Power Unit.
WARNING: NEVER CLEAN THE FILTER WITH COMPRESSED AIR OR A BRUSH. HEPA FILTERS ARE EASILY DAMAGED BY THE USE
OF COMPRESSED AIR OR BY BRUSHING.
Misuse of CleanSpace Respirator by cleaning the filter may result in overexposure to contaminants and lead to sickness or
death.
•
Wipe down the outside of the Power Unit using CleanSpace Cleaning Wipes or a damp cloth.
•
Allow the Power Unit to air dry in a clean environment. The Power Unit can also be hand-dried with a clean lint-free cloth.
•
Note: Do not dry the Power Unit by exposing it directly to heat (e.g. hair dryers or heaters) or compressed air.
CLEANING THE HARNESS AND NECK SUPPORT
•
The harness and neck support can be washed in warm soapy water or machine washed. After cleaning, allow the harness
and neck supports to air dry in a clean environment.
•
Note: Do not dry the Power Unit by exposing it directly to heat (e.g. hair dryers or heaters).
15. Recalibration
You must recalibrate the internal pressure sensor any time that the CleanSpace Respirator is exposed to changes in
temperature of more than 20°C.
Recalibration procedure:
1. With no mask attached, place the Power Unit on a flat surface with the keypad facing up.
2. With the Power Unit in standby mode (one or more Battery Indicator Light(s) illuminated), press and hold both the Power
Button and the FLOW TEST Button.
3. When both the blue and red indicator lights illuminate, release both buttons.
4. Do not move the Power until the red and blue indicator lights extinguish; approximately(10 – 15 seconds).
5. Recalibration is complete.
The CleanSpace Respirator automatically recalibrates the system any time it is left undisturbed in Standby Mode for more
than 20 seconds.
16. Approvals
CleanSpace Respirators are a complete respiratory protection system. In addition, some combinations of components have
been approved for use in some explosive atmospheres.
Only specific combinations of components have been approved for use in explosive atmospheres.
The IECEx / ATEX / ETL approvals apply only when all components are IECEx / ATEX / ETL approved. If the equipment is to be
used in an explosive atmosphere, ensure proper approval exists when purchasing spare parts and accessories.
Not all components and accessories shown in this manual carry IECEx / ATEX / ETL approval.
SEE BELOW FOR A LIST OF THOSE COMPONENTS THAT HAVE IECEx / ATEX APPROVAL