can be cancelled by pressing down the OFF button, then pressing the ON
button, and then releasing the OFF button after the LEDs rapidly flash amber.
5. If the study you are scheduling will be done at a location in a time zone different from
the one your computer is set to, check the checkbox labeled “Study will take place in
a different time zone” and select the time zone using the drop-down list provided.
This will adjust the internal clock within the SleepView to the time zone where the
study will be performed.
6. Configure the auxiliary port for the appropriate sensor that will be used for the study.
If no sensor will be used, select ”(None)” to disable the LED for this channel only. It
will remain off during use.
IMPORTANT NOTE: The next step will cause all previously recorded studies to
be deleted from the device.
7. Click <Next> and the “Downloading Parameters” page will appear and begin to send
all of the options you selected to the device.
8. Once programming is complete, all of the operations will report “DONE” next to its
label and the <NEXT> button will become active. Click <Next> and the “Insert
Battery and Check Device” page will appear.
PLEASE NOTE: If you did not program the SleepView Monitor to automatically
turn ON then “N/A” will appear next to “Setting auto-start time”
9. The SleepView will turn OFF after programming is complete.
IMPORTANT NOTE: The next steps must be performed in order to commit the
settings to the device. You must comply within 2 minutes to ensure proper
configuration.
10. Click <NEXT> and the “Insert Battery and Check Device” page will appear.
Disconnect the SleepView from the USB cable and within 2 MINUTES insert a new
AAA battery.
11. Press and hold the ON button until the power LED lights green.
IMPORTANT NOTE: If the power LED does not illuminate green and flashes
amber, then the settings did not properly transfer into the device.