24
2. Press to mark this test as a control solution test
With “CTL” displayed, the meter will not store your test result in
memory. If you press
again, the “CTL” will disappear and this test
is no longer a control solution test.
When doing the control solution test, you have to mark it so that
the test result will not be stored in the memory. Failure to do so
will mix up the blood glucose test results with the control solution
test results in memory.
WARNING
Performing a Control Solution Test
1. Insert the test strip to turn on the meter
Insert the test strip into the meter. Wait for the meter to display the
test strip and blood drop symbols.